Protein Concentration Calculator (A280)
Turns a UV absorbance reading (usually A280) into a protein concentration with the Beer–Lambert law. Pick a common protein such as BSA, IgG or lysozyme, or enter your own extinction coefficient as a molar value (M⁻¹cm⁻¹), as the absorbance of a 1 mg/mL solution, or as E1%. Add the cuvette path length in cm or mm and any dilution factor, and get the concentration in mg/mL, µg/mL and molar units.
- The Beer–Lambert law holds, which is true for dilute, clear solutions within the instrument's linear range.
- The absorbance has already been corrected with a buffer blank; light scattering by aggregates is not corrected.
- Preset extinction coefficients and molecular weights are published or sequence-calculated reference values and may differ from your exact construct, tag or glycoform.
- All absorbance is assumed to come from the protein; nucleic acids or other UV-absorbing compounds will bias the result.
Formula
Beer–Lambert law: A = ε × c × b, so c (mol/L) = A × n ÷ (ε × b) and c (mg/mL) = c (mol/L) × MW. With a mass coefficient: c (mg/mL) = A × n ÷ (Abs 0.1% × b) = 10 × A × n ÷ (E1% × b).
- A is the blank-corrected absorbance (no unit), ε the molar extinction coefficient in M⁻¹cm⁻¹, b the path length in cm, n the dilution factor and MW the molecular weight in g/mol.
- Because g/L equals mg/mL, multiplying the molar concentration by the molecular weight gives mg/mL directly.
- Abs 0.1% is the absorbance of a 1 mg/mL solution in a 1 cm cell, equal to ε ÷ MW. E1% is the absorbance of a 1% (10 mg/mL) solution, so E1% = 10 × Abs 0.1%.
- A path length in mm is divided by 10 to give cm. Microvolume instruments often report a value already normalised to 1 cm.
- Preset ε and MW values come from the sources listed below: IgG and BSA from Thermo Fisher's extinction-coefficient tech tip, APC and PE from Thermo Fisher's Molecular Probes handbook, ATP from Thermo Fisher's ATP product guide, Trp, Tyr and cystine from Gill and von Hippel's values for unfolded protein, phenylalanine from the PhotochemCAD spectral database, and the other proteins from ExPASy ProtParam. Some presets are rounded or differ slightly from these sources: APC MW 105,000 (handbook 104,000), BSA MW 66,463 (66,400 in the tech tip, 66,433 from the sequence), phenylalanine ε 200 (195 at 257.5 nm), lysozyme ε 37,901 and MW 14,000 (ProtParam 37,970 and 14,313), and streptavidin ε 176,000 and MW 55,000 for the tetramer (ProtParam gives about 167,760 and 53,086 for four core subunits). The cysteine preset's ε of 120 is the value per cystine (disulfide), while its MW is that of free cysteine.
- APC, PE, ATP and phenylalanine absorb most strongly away from 280 nm (650, 565, 259 and 257 nm), so read the absorbance at that wavelength for them.
IgG, A280 = 0.70, 1 cm cuvette, undiluted
Inputs
- Protein: igg
- Absorbance (A280): 0.7
- Path length (b): 1
- Path length unit: cm
- Dilution factor (n): 1
IgG has ε = 210,000 M⁻¹cm⁻¹ and MW = 150,000 g/mol. Molar concentration = 0.70 ÷ (210,000 × 1) = 3.333 × 10⁻⁶ M = 3.333 µM. Mass concentration = 3.333 × 10⁻⁶ × 150,000 = 0.5 g/L = 0.5 mg/mL (500 µg/mL). The matching Abs 0.1% is 210,000 ÷ 150,000 = 1.4.
Sources & references
- Extinction Coefficients (Tech Tip TR0006)
- How to measure and predict the molar absorption coefficient of a protein (Pace et al., Protein Science 4:2411–2423, 1995)
- ProtParam: rabbit fructose-bisphosphate aldolase A (UniProt P00883, chain 2–364)
- ProtParam: bovine pancreatic ribonuclease A (UniProt P61823, chain 27–150)
- ProtParam: bovine insulin B chain (UniProt P01317, 25–54)
- ProtParam: bovine insulin A chain (UniProt P01317, 85–105)
- Molecular Probes Handbook, Table 6.2: Spectral data for B-PE, R-PE and APC
- Thermo Scientific ATP Solution (R0441) user guide
- Measuring Protein Concentration (Biophysics Facility protocol, T. Sharpe, 2012)
- PhotochemCAD spectrum: Phenylalanine in water
- ProtParam: bovine serum albumin (UniProt P02769, chain 25–607)
- ProtParam: hen egg-white lysozyme (UniProt P00698, chain 19–147)
- ProtParam: core streptavidin (UniProt P22629, residues 37–163)
Frequently asked questions
Why measure protein at 280 nm?
The aromatic side chains of tryptophan and tyrosine absorb UV light strongly near 280 nm, and disulfide bonds (cystines) add a little. Most proteins contain these residues, so A280 is a quick, reagent-free way to estimate concentration. Proteins without Trp or Tyr barely absorb at 280 nm and need another method such as a Bradford or BCA assay.
What is the difference between molar ε, Abs 0.1% and E1%?
They describe the same property on different scales. Molar ε is the absorbance of a 1 mol/L solution over 1 cm. Abs 0.1% is the absorbance of a 1 mg/mL (0.1%) solution, and E1% that of a 10 mg/mL (1%) solution. Convert with Abs 0.1% = ε ÷ MW and E1% = 10 × ε ÷ MW. For IgG, ε = 210,000 and MW = 150,000 give Abs 0.1% = 1.4 and E1% = 14 (13.7 is also widely quoted).
Where do I get the extinction coefficient of my protein?
Estimate it from the amino-acid sequence: ε280 ≈ 5,500 × (number of Trp) + 1,490 × (number of Tyr) + 125 × (number of cystines), the Pace method used by ProtParam. The result is usually within about 5% of the measured value for folded proteins. For an unknown mixture, an Abs 0.1% of 1.0 is a common rough assumption.
What absorbance range gives reliable results?
Most cuvette spectrophotometers are linear up to about 1 to 2 absorbance units, and readings below about 0.1 are dominated by noise. If A is above that range, dilute the sample, measure again and enter the dilution factor. Always subtract a blank of the same buffer.
How does the dilution factor work?
If you mixed 10 µL of sample with 90 µL of buffer, the dilution factor is 100 ÷ 10 = 10. The calculator works out the concentration in the cuvette and multiplies it by the dilution factor to give the concentration of the original, undiluted sample.
Can nucleic acids interfere with A280?
Yes. DNA and RNA absorb strongly at 260 nm and also at 280 nm, so contamination inflates the protein estimate. An A260/A280 ratio near 0.57 suggests pure protein; higher ratios point to nucleic acid.
Related calculators
Protein Molecular Weight Calculator Works out the molecular weight of a peptide or protein from its amino-acid sequence in one-letter code. It adds up the masses of the amino acids and removes one water molecule for every peptide bond, giving the result in daltons, kilodaltons and g/mol, as an average or monoisotopic mass. It also lists the amino-acid composition, the three-letter sequence and an estimated extinction coefficient at 280 nm. DNA Concentration Calculator Turns a UV spectrophotometer reading into a nucleic acid concentration. Enter the absorbance at 260 nm, the cuvette path length and any dilution factor, and pick double-stranded DNA (50 µg/mL per A₂₆₀), single-stranded DNA (33), RNA (40), your own conversion factor, or a custom oligonucleotide. For an oligo, paste its sequence and the calculator works out the molecular weight and the nearest-neighbour extinction coefficient, or type both values in. Optional A₂₈₀ and A₂₃₀ readings give the purity ratios, and a sample volume gives the total yield. Molarity Calculator Calculates the molarity of a solution, or works backwards to the mass of solute or the volume of solution you need for a target molarity. Molarity is the amount of dissolved substance in moles per litre of solution (mol/L, written M). Enter the molar mass with the mass of solute and the final volume, a target molarity, or a mass concentration in g/L, and get the molarity, moles of solute and concentration. Cell Dilution Calculator Plan a cell suspension dilution with C1 × V1 = C2 × V2. Enter any three of the stock concentration, the volume of stock to take, the target concentration and the final volume, each with its own unit (µL, mL, cL or L), and the calculator finds the fourth. It also reports the dilution factor, how much medium or buffer to add, the total number of cells, and a serial dilution plan when the stock volume is too small to pipette accurately.